FDA Adverse Event Malfunction Summary report: N

INOMAX DSIR (DELIVERY SYSTEM)

MDR report key: 4173553 · Received August 7, 2014

Report

Report Number
3004531588-2014-00041
Event Type
Malfunction
Date Received
August 7, 2014
Report Date
July 11, 2014
Manufacturer
IKARIA
Product Code
MRN
PMA / PMN Number
K061901
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY. INOMAX DSIR# (B)(4) WAS AT THE REGIONAL SERVICE CENTER (RSC) IN (B)(6) FOR SERVICE. DURING POST-SERVICE BURN IN, INOMAX DSIR# (B)(4) EXPERIENCED A DELIVERY FAILURE (DF) ALARM. A REVIEW OF THE POST-SERVICE LOG CONFIRMED A DF ALARM DUE TO BACKLIGHT CURRENT BELOW MINIMUM (ACTUAL: 793 COUNTS; MINIMUM: 800 COUNTS). THE RSC INVESTIGATION DETERMINED THAT THE ROOT CAUSE OF THE ISSUE WAS A FAULTY 50018 MAIN BOARD AND REPLACED IT TO RECTIFY THE FAULT. A FULL FUNCTIONAL TEST WAS PERFORMED AND THE DEVICE OPERATED ACCORDING TO SPECIFICATIONS SO IT WAS RETURNED TO THE DEVICE SERVICE POOL. THE ROOT CAUSE WAS A DISPLAY-BACKLIGHT FAILURE. THIS CONDITION WILL BE TRACKED AND TRENDED UNDER (B)(6) QUALITY SYSTEM. THIS DEVICE WAS NOT IN USE ON A PATIENT; HOWEVER, IT IS BEING SUBMITTED TO REGULATORY AUTHORITIES BECAUSE A SIMILAR FAILURE OCCURRED IN THE PAST WHICH RESULTED IN A SERIOUS ADVERSE EVENT (MDR 3004531588-2013-00023).

Description of Event or Problem · 1

DISPLAY BACKLIGHT FAILURE [DEVICE ISSUE]. CASE DESCRIPTION: ON (B)(6) 2014, DURING A ROUTINE REVIEW OF SERVICE ORDER FINDINGS FROM (B)(6) REGIONAL SERVICE CENTER (RSC) IN (B)(6) , AN (B)(6) QUALITY MANAGER IDENTIFIED THAT INOMAX DSIR# (B)(4) HAD EXPERIENCED A DISPLAY BACKLIGHT FAILURE DURING POST-SERVICE BURN IN. THIS DEVICE ISSUE IS SIMILAR TO A DEVICE ISSUE WHICH OCCURRED IN THE PAST AND HAD RESULTED IN A SERVICE ADVERSE EVENT. DURING POST-SERVICE BURN IN, INOMAX DSIR# (B)(4) EXPERIENCED A DELIVERY FAILURE (DF) ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463143 INOMAX DSIR (DELIVERY SYSTEM) APPARATUS, NITRIC OXIDE DELIVERY MRN IKARIA 10007

Patients

Seq Age Sex Outcome Treatment
1