FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUAL CHANNEL

MDR report key: 4173530 · Received August 7, 2014

Report

Report Number
9615050-2014-04715
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
July 24, 2014
Report Date
July 24, 2014
Manufacturer
HOSPIRA HOLDINGS DE COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0070-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE TOUCHSCREEN DID NOT RESPOND ALONG THE RIGHT SIDE OF THE DISPLAY. THE PROBABLE CAUSE WAS DUE TO CONTAMINATION ON THE BOTTOM OF THE TOUCHSCREEN DUE TO FLUID INGRESS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN DOES NOT RESPOND WHEN PRESSED. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463166 SYMBIQ DUAL CHANNEL 80 FRN HOSPIRA HOLDINGS DE COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1