FDA Adverse Event Malfunction Summary report: N

SYMBIQ 3.13 DUAL CHA

MDR report key: 4173509 · Received October 10, 2014

Report

Report Number
9615050-2014-05577
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
January 1, 2014
Report Date
September 25, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0070-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING THE DEVICE DID NOT PASS THE TOUCHSCREEN TEST AND WAS OUT OF CALIBRATION. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH A NOTE THAT STATED, "NO RESPONSE FROM TOUCHSCREEN." NO INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DID NOT PASS THE TOUCHSCREEN TEST AND THE TOUCHSCREEN WAS OUT OF CALIBRATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
642847 SYMBIQ 3.13 DUAL CHA 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK