HOMECHOICE
Report
- Report Number
- 1416980-2014-35926
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Date of Event
- July 26, 2014
- Report Date
- September 20, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS USE ERROR, INAPPROPRIATE BYPASS OF THE DRAIN, NOT INCLUDING INITIAL DRAIN. HOMECHOICE / HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE GIVES INSTRUCTIONS ON HOW TO BYPASS THE DRAIN PHASE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 22:00:00. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 2344ML, INDICATING THE HOME PATIENT (HP) DRAINED 2344ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654859 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |