ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-29286
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Report Date
- October 2, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 12/02/2014 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/11/2014 WITH THE FOLLOWING FINDINGS: THE ORIGINAL COMPLAINT COULD NOT BE DUPLICATED. THERE WERE MULTIPLE LOSS OF PRIME WARNINGS DUE TO LOW, NON-ZERO FORCE OBSERVED IN THE BLACK BOX HISTORY. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO LOSS OF PRIME WARNINGS OR OTHER ALARMS OBSERVED. THE FORCE SENSOR CALIBRATION WAS FOUND TO BE WITHIN SPECIFICATION.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.