NOVASURE IMPEDANCE CONTROLLED EA SYSTEM
Report
- Report Number
- 1222780-2014-00163
- Event Type
- Injury
- Date Received
- September 19, 2014
- Date of Event
- August 1, 2014
- Report Date
- August 20, 2014
- Manufacturer
- HOLOGIC
- Product Code
- MNB
- PMA / PMN Number
- P010013
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
CONCOMITANT PRODUCT: SERIAL NUMBER OF THE RADIO FREQUENCY CONTROLLER NOT PROVIDED BY THE COMPLAINANT. THE DEVICE HAS NOT YET BEEN RETURNED, THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. IF THE DEVICE IS RETURNED AND EVALUATION COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE REPORTED LOT NUMBER. THE LOT WAS RELEASED MEETING ALL QA SPECIFICATIONS. CURRENTLY UNABLE TO ESTABLISH A RELATIONSHIP OR IMPACT TO THE REPORTED OBSERVATION. (B)(4).
IT WAS REPORTED A PHYSICIAN ATTEMPTED A NOVASURE ENDOMETRIAL ABLATION AND HAD DIFFICULTY SEATING THE DEVICE. ONCE THE DEVICE WAS SEATED, THE PHYSICIAN RECEIVED SEVERAL CAVITY INTEGRITY ASSESSMENT (CIA) TEST FAILURES. A HYSTEROSCOPY WAS PERFORMED AND A UTERINE PERFORATION WAS VISUALIZED. THE PROCEDURE WAS ABORTED AND NO INTERVENTION WAS REQUIRED. THE PATIENT WAS DISCHARGED HOME. A HYSTEROSCOPY, DILATION, AND SOUNDING WITH A METAL SOUND (NOT HOLOGIC DEVICES) WERE PERFORMED PRIOR TO THE ATTEMPTED ABLATION. IT IS NOT KNOWN WHEN THIS PERFORATION OCCURRED OR WHAT INSTRUMENT MAY HAVE BEEN THE CAUSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 583014 | NOVASURE IMPEDANCE CONTROLLED EA SYSTEM | UTERINE ABLATION DEVICE | MNB | HOLOGIC | NS2000 | 14306RD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other | RADIO FREQUENCY CONTROLLER - SERIAL NUMBER UNKNOWN |