FDA Adverse Event
Malfunction
Summary report: N
ANSPACH BLACK MAX SYSTEM
MDR report key: 4173101
·
Received October 15, 2014
Report
- Report Number
- 1045834-2014-13860
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Report Date
- August 14, 2012
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- ERL
- PMA / PMN Number
- PK831756
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED DAMAGED COMPONENTS. THIS IS ATTRIBUTED TO "ACESSIVE" WEAR. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER ON (B)(4) 2012.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE DEVICE DID NOT HAVE ENOUGH POWER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654083 | ANSPACH BLACK MAX SYSTEM | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |