FDA Adverse Event Malfunction Summary report: N

11 CM ANGLE ATTACHMENT QD11

MDR report key: 4173099 · Received October 15, 2014

Report

Report Number
1045834-2014-14437
Event Type
Malfunction
Date Received
October 15, 2014
Report Date
May 10, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED DEFECTIVE COMPONENTS. THIS IS ATTRIBUTED TO FAULTY MAINTENANCE. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

DEVICE WAS RETURNED FOR REPAIR WITH DAMAGED BEARINGS AND GEARS. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
655780 11 CM ANGLE ATTACHMENT QD11 MOTOR, DRILL, ELECTRIC HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1