ANSPACH® XMAX
Report
- Report Number
- 1045834-2014-14562
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Report Date
- January 18, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- ERL
- PMA / PMN Number
- PK131053
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE DISCS WERE WORN, THE BLADES WERE DAMAGED, THE DEVICE WAS TOO LOUD, THE COLORING WAS MISSING FROM THE COUPLING PIECE, AND THE HOSE HAD A TEAR. THIS WAS ATTRIBUTED TO NORMAL WEAR. THE HOSE, DISCS, BEARING, SHAFT, BOLTING, STATOR, AND COUPLING SLEEVE WERE REPLACED. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, TECHNICIAN FOUND DISCS WORN DOWN, TOO LOUD AND SLOW, HOSE HAS A TEAR AND RED COLORING IS MISSING FROM THE COUPLING PLACE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 653947 | ANSPACH® XMAX | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |