FDA Adverse Event
Malfunction
Summary report: N
TERUMO TCM II COOLING AND HEATING SYSTEM
MDR report key: 4172361
·
Received October 9, 2014
Report
- Report Number
- 1828100-2014-00857
- Event Type
- Malfunction
- Date Received
- October 9, 2014
- Date of Event
- September 18, 2014
- Report Date
- September 18, 2014
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- K883603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED COMPLAINT WAS NOT CONFIRMED. THE FIELD SVC REP (FSR) RAN THE DEVICE FOR ONE HR WITHOUT ANY FAILURES. A CALIBRATION CHECK WAS COMPLETED SUCCESSFULLY WITHOUT HAVING TO MAKE ANY ADJUSTMENTS. IF ADD'L INFO BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE COOLER HEATER UNIT WOULD REVERT TO "STANDBY" MODE WHILE IN THE MAINTAIN MODE AND "REWARM" MODE. THE UNIT NEVER FAULTED DURING THE SURGICAL PROCEDURE. THE DEVICE WAS NOT CHANGED OUT. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, AND NO ADVERSE CONSEQUENCES TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 636788 | TERUMO TCM II COOLING AND HEATING SYSTEM | TCM II COOLING AND HEATING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 4415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |