FDA Adverse Event Malfunction Summary report: N

HOMEPUMP C-SERIES 270ML, 10 ML/HR

MDR report key: 4172264 · Received October 9, 2014

Report

Report Number
2026095-2014-00195
Event Type
Malfunction
Date Received
October 9, 2014
Date of Event
June 30, 2014
Report Date
September 12, 2014
Manufacturer
I-FLOW, LLC
Product Code
MEB
PMA / PMN Number
K052117
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

METHOD: THE DEVICE WERE REPORTED NOT TO BE AVAILABLE AS THEY WERE DISCARDED, THEREFORE TESTING METHODS COULD NOT BE PERFORMED. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS CONDUCTED FOR THE LOT NUMBER REPORTED. RESULTS: ACCORDING TO THE DHR REVIEW, THE PRODUCTION LOT MET ALL MANUFACTURING AND QUALITY SPECIFICATIONS. AS A DEVICE WAS NOT AVAILABLE FOR AN EVAL NO METHODS WERE PERFORMED, THEREFORE, RESULTS CANNOT BE OBTAINED. CONCLUSIONS: THE DEVICE WAS NOT RETURNED TO I-FLOW FOR EVAL, THEREFORE, WE ARE UNABLE TO DETERMINE AN EXACT CAUSE FOR THE REPORTED EVENT. BASED ON THE REPORTED INFO, THE DEVICE WAS USED LESS THAN ONE MINUTE AFTER FILLING. PRODUCT COMPLAINT AND MDR TREND REPORTING SYSTEMS. TREND INFO IS USED TO IDENTIFY THE NEED FOR ADD'L INVESTIGATIONS.

Description of Event or Problem · 1

FILL VOLUME: EXACT AMOUNT UNK, BETWEEN 140ML OR 160 ML.. FLOW RATE: 10 ML/HR. {PLEASE REFERENCE 2026095-2014-00138/14-00589 (A), 2026095-2014-00139/14-00589 (B) AND 2026095-2014-00140/14-00589 (C) AND 2026095-2014-00141/14-00899 (D) WHICH HAVE ALREADY BEEN REPORTED INITIALLY ON 08/13/2014}. ADD'L INFO WAS PROVIDED REGARDING PT #2 AND TWO ADD'L FAST FLOW INCIDENTS WERE REPORTED FOR THIS PT. PLEASE REFERENCE 2026095-2014-00196/14-00589 (F) WHICH IS BEING REPORTED ALONG WITH THIS REPORT. PT 2, INCIDENT 1 OF 2: BEGINNING ON (B)(6) 2014 FOR 15 DAYS, INFUSIONS HAVE NOT BEEN RUNNING AS EXPECTED. TWO ADD'L FAST FLOW INCIDENTS WERE REPORTED FOR PT TWO. IT WAS WAITED LESS THAN ONE MINUTE BETWEEN FILLING THE PUMP AND PLACING IT ON THE PT. IT WAS REPORTED THAT THE INFUSION TIME WAS WITHIN 11 HRS, INSTEAD OF 12 HRS AND THIS OCCURRED TWICE. THE PUMP WAS FILLED WITH TAZOCILLINE/NACI 0.9%. THE PT WAS REPORTED TO HAVE EXPERIENCED HYPERTHERMY ON DAY 1 AND DAY 2 OF THE TREATMENT AND WAS TREATED WITH PARACETAMOL. NO FURTHER INFO REGARDING THIS PT WAS REPORTED. THE INDICATION OF THE ANTIBIOTIC THERAPY WAS UNK. THE PUMPS HAVE BEEN DISCARDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
636838 HOMEPUMP C-SERIES 270ML, 10 ML/HR ELASTOMERIC PUMP MEB I-FLOW, LLC C270100 201122658

Patients

Seq Age Sex Outcome Treatment
1 25 YR TAZOCILLINE/NACI 0.9%