FDA Adverse Event Malfunction Summary report: N

SYMBIQ DCHANNEL 3.01

MDR report key: 4172257 · Received October 9, 2014

Report

Report Number
9615050-2014-05542
Event Type
Malfunction
Date Received
October 9, 2014
Date of Event
January 1, 2014
Report Date
September 23, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0070-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THAT THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. THE PROBABLE CAUSE WAS BROKEN TOUCHSCREEN. THIS WAS DUE TO CONTAMINATION CAUSED BY FLUID INGRESS WHICH ALTERED THE CHARACTERISTICS OF THE TOUCHSCREEN. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN WOULD NOT CALIBRATE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN WOULD NOT CALIBRATE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
636983 SYMBIQ DCHANNEL 3.01 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK