FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4172246 · Received October 15, 2014

Report

Report Number
2531779-2014-29251
Event Type
Malfunction
Date Received
October 15, 2014
Report Date
October 3, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE KEYPAD COVER WAS INTACT. EVALUATION REVEALED THAT ALL OF THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE: THE REPORTED ISSUE WAS DUPLICATED. THE KEYPAD COVER WAS REMOVED, AND EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL OF THE KEY CONTACTS; THIS DEVICE FAILURE DIRECTLY CAUSED THE REPORTED ISSUE. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE UP ARROW, DOWN ARROW, AND OK KEYPAD BUTTONS WERE UNDER-RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.