ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-29249
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Report Date
- October 3, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE KEYPAD COVER WAS INTACT. EVALUATION REVEALED THAT ALL OF THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE: THE REPORTED ISSUE WAS DUPLICATED. THE KEYPAD COVER WAS REMOVED, AND EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL OF THE KEY CONTACTS; THIS DEVICE FAILURE DIRECTLY CAUSED THE REPORTED ISSUE. UNRELATED TO THE REPORTED ISSUE, THE DISPLAY SCREEN VISUAL WAS FOUND TO BE DISCOLORED DUE TO AN UNIDENTIFIED DISPLAY FAILURE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE UP AND DOWN ARROW KEYPAD BUTTONS WERE UNDER-RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654871 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |