FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4171922 · Received October 15, 2014

Report

Report Number
1031452-2014-13928
Event Type
Malfunction
Date Received
October 15, 2014
Report Date
September 25, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

PER THE PROVIDER, THE UNIT IS ALARMING AND DISPLAYING A LOW O2 ERROR. INDICATION IS THAT SIEVE BED FILTERS (BOTTOM) ARE NOT SEATED PROPERLY ALLOWING SIEVE MATERIAL TO BECOME PULVERIZED AND BYPASS FILTER. SIEVE DUST IS MIGRATING INTO THE MANIFOLD VALVE AND MUFFLER. REFER TO CAPA (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
655910 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other