OT PING METER
Report
- Report Number
- 2939301-2014-27514
- Event Type
- Malfunction
- Date Received
- October 15, 2014
- Report Date
- October 7, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K082590
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 ¿ (12/02/2014). THE PATIENT¿S METER HAS BEEN RETURNED ON 10/17/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 11/18/2014 WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED; HOWEVER, A SECONDARY ISSUE WAS NOTED WHEN THE METER WAS FOUND TO HAVE DIRTY SPC PINS. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING DOES NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 654005 | OT PING METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3352187 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |