FDA Adverse Event Injury Summary report: N

OT ULTRA2 METER

MDR report key: 4171825 · Received October 15, 2014

Report

Report Number
2939301-2014-27474
Event Type
Injury
Date Received
October 15, 2014
Date of Event
September 28, 2014
Report Date
October 7, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K053529
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

ON (B)(6) 2014, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) ALLEGING THAT HER ONETOUCH ULTRA2 METER WAS READING INACCURATELY HIGH. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE¿S (CCA) DOCUMENTATION. THE PATIENT CLAIMED THE ALLEGED ISSUE BEGAN ON (B)(6) 2014 IN THE AFTERNOON. SHE TESTED ON THE SUBJECT METER AND OBSERVED A VALUE OF ¿157MG/DL¿ WHICH SHE FELT WAS INACCURATELY HIGH AS COMPARED TO HER USUAL READINGS/FEELINGS. IT IS NOT KNOWN WHAT RANGE THE PATIENT CONSIDERS TO BE NORMAL. THE PATIENT DOES NOT TAKE ANY MEDICATIONS TO MANAGE HER DIABETES. IT IS NOT KNOWN WHAT ACTION THE PATIENT TOOK IN RESPONSE TO THE ALLEGED HIGH READING. THE PATIENT CLAIMED THAT IMMEDIATELY AFTER THE ALLEGED ISSUE OCCURRED, THE PATIENT DEVELOPED SYMPTOMS OF ¿SHAKES, DIZZINESS AND LIGHT HEADEDNESS¿ BUT DESPITE HER SYMPTOMS, SHE DENIED RECEIVING ANY MEDICAL TREATMENT. THE PATIENT REPORTED THAT SHE PURCHASED A NEW METER IMMEDIATELY. AT THE TIME OF TROUBLESHOOTING THE CCA CONFIRMED THE METER WAS SET TO THE CORRECT UNIT OF MEASURE, THE PATIENT DID NOT HAVE ANY MORE TEST STRIPS AVAILABLE FOR TROUBLESHOOTING. REPLACEMENT PRODUCTS HAVE BEEN SENT TO THE PATIENT AND SUBJECT PRODUCTS ARE PENDING RETURN FOR INVESTIGATION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SYMPTOMS SUGGESTIVE OF HYPOGLYCEMIA AFTER OBTAINING AN INACCURATE HIGH BLOOD GLUCOSE RESULT ON THE SUBJECT METER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
653837 OT ULTRA2 METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3642388

Patients

Seq Age Sex Outcome Treatment
1 44 YR Life Threatening