FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4171735 · Received October 14, 2014

Report

Report Number
2531779-2014-29230
Event Type
Malfunction
Date Received
October 14, 2014
Report Date
October 2, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: ON (B)(6) 2014 BLACK BOX RECORDED MULTIPLE ¿LOSS OF PRIME¿ WARNINGS DUE TO LOW, NON-ZERO, FORCE SENSOR READINGS. THE PUMP WAS EXERCISED FOR 24 HOURS AT 1 UNIT/HOUR WITH NO ¿LOSS OF PRIME¿ WARNINGS OR OTHER EAWS OBSERVED. INVESTIGATORS CHECKED FORCE SENSOR CALIBRATION AND THE PUMP WAS FOUND TO BE WITHIN CALIBRATION AT 173. UNABLE TO DUPLICATE CUSTOMER COMPLAINT. RETURNED BATTERY CAP USED FOR INVESTIGATION.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
650008 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 11 YR