ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-29229
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Report Date
- October 2, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 12/22/2014 - DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/03/2014 WITH THE FOLLOWING FINDINGS: THE PUMP POWERED ON WITH THE RETURNED BATTERY CAP. THERE WERE MULTIPLE 064 MOTOR ERRORS OBSERVED IN THE PUMP'S BLACK BOX ON (B)(4) 2014. THE ERRORS OCCURRED DURING THE "LOAD" SEQUENCE. DURING INVESTIGATION, A HARD 064 MOTOR ERROR WAS RECEIVED ON EACH "LOAD" ATTEMPT. INVESTIGATION COULD NOT BE COMPLETED AS A RESULT OF THIS ALARM. THERE WAS NO EVIDENCE OF MOISTURE INSIDE THE PUMP. FURTHER INVESTIGATION SHOWED AN INTERNAL MOTOR FAILURE CAUSING THE 064 MOTOR ERRORS. THE MOTOR WAS REMOVED FROM PUMP AND THE FAILURE WAS DUPLICATED ON A TEST FIXTURE. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE PUMP EMITTED A CALL SERVICE 064 ALARM. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 650331 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |