FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4171715 · Received October 14, 2014

Report

Report Number
2531779-2014-29210
Event Type
Malfunction
Date Received
October 14, 2014
Report Date
October 1, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/19/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWS NO EVIDENCE OF A LOAD STEP MALFUNCTION. NO DAMAGE OBSERVED TO PISTON CAP. NO DEBRIS OBSERVED IN CARTRIDGE COMPARTMENT. A REWIND, LOAD AND PRIME SEQUENCE WERE PERFORMED SUCCESSFULLY WITH NO ISSUES. A FORCE SENSOR CALIBRATION CHECK SHOWED THE PUMP IS DETECTING THE CORRECT FORCE AT 5 LBS. INVESTIGATORS WERE UNABLE TO CONFIRM OR DUPLICATE THE COMPLAINT. THERE WAS NO DEFECT FOUND.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOAD STEP MALFUNCTION) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
650093 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 59 YR