FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 4171709 · Received October 14, 2014

Report

Report Number
2531779-2014-29203
Event Type
Malfunction
Date Received
October 14, 2014
Report Date
October 1, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE CARTRIDGE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/13/2014 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A FORCE TEST AND FILL TEST WERE PERFORMED WITH NO FAILURES OBSERVED. THE CARTRIDGE FORCE READINGS WERE CONFIRMED TO BE WITHIN SPECIFICATIONS; NO FAILURES WERE NOTED RELATED TO THE LOSS OF PRIME COMPLAINT. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: A RETAIN SAMPLE FROM THE SAME LOT NUMBER WAS TESTED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A FILL TEST WAS PERFORMED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO LEAKS OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. THE CARTRIDGE FORCE READINGS WERE CONFIRMED TO BE WITHIN SPECIFICATIONS; NO FAILURES WERE NOTED RELATED TO THE LOSS OF PRIME COMPLAINT. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE, CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION. THE RETAIN CARTRIDGE PASSED A LOT REVIEW.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE SAME CARTRIDGE CAUSED MULTIPLE LOSS OF PRIME WARNINGS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
650158 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION D200088

Patients

Seq Age Sex Outcome Treatment
1 62 YR