FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4171696 · Received October 14, 2014

Report

Report Number
2531779-2014-29196
Event Type
Malfunction
Date Received
October 14, 2014
Report Date
September 30, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 01/23/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 01/09/2015 WITH THE FOLLOWING FINDINGS: THE BLACK BOX AND ALARM HISTORY SHOWS NO ACTIVITY OUTSIDE OF NORMAL USE. THE RETURNED BATTERY CAP AND CARTRIDGE CAP WERE USED TO COMPLETE THE INVESTIGATION. NO EAW OCCURRED DURING THE INVESTIGATION, THE PRODUCT PERFORMS WITHIN SPECIFICATIONS. INVESTIGATORS WERE UNABLE TO DUPLICATE ¿UNUSUAL ISSUE¿ COMPLAINT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2014 ALLEGING THAT THEY ARE HAVING PUMP ISSUES. THERE IS NO ADDITIONAL INFORMATION AT THE TIME OF THIS REPORT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
650148 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 19 YR