MEGA NEEDLE DRIVER INSTRUMENT

MDR report key: 4171609 · Received October 14, 2014

Report

Report Number
2955842-2014-05481
Event Type
Malfunction
Date Received
October 14, 2014
Date of Event
August 4, 2014
Report Date
September 16, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION REVEALED THE PITCH UP CABLE BROKE OFF AND WAS MISSING FROM THE DISTAL END. THE CRIMP THAT CONTAINS THE CABLE WAS MISSING. NO OTHER DAMAGE FOUND. ISI HAS CONDUCTED A DEVICE HISTORY RECORD (DHR) REVIEW FOR THIS DEVICE AND DID NOT FIND ANY NON-CONFORMANCES THAT WERE RELATED TO THIS REPORTED EVENT. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE BROKEN PITCH CABLE, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY PROCEDURE, A WIRE WAS STICKING OUT OF THE MEGA NEEDLE DRIVER INSTRUMENT. THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.