FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4171456 · Received October 14, 2014

Report

Report Number
2032227-2014-37581
Event Type
Malfunction
Date Received
October 14, 2014
Date of Event
September 7, 2014
Report Date
September 14, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TESTS. NO MOTOR ERROR OR EXCESSIVE NO DELIVERY ALARMS WERE NOTED. THE MOTOR PASSED MOTOR TEST. THE INSULIN PUMP WAS MONITORED FOR SEVERAL DAYS AND NO BLANK DISPLAY WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH NORMAL OPERATING CURRENTS. NO DAMAGE WAS FOUND ON THE CONNECTOR LCD ISOLATION TAPE NOTED. NO UNEXPECTED LOW BATTERY ALARM NOTED. THE INSULIN PUMP WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW, MISSING END CAP STICKER AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM AND NO DELIVERY ALARMS. TROUBLESHOOTING OCCURED. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
650563 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAS

Patients

Seq Age Sex Outcome Treatment
1 69 YR