FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4171308 · Received October 14, 2014

Report

Report Number
1031452-2014-13866
Event Type
Malfunction
Date Received
October 14, 2014
Report Date
September 25, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

ALARMING WITH LOW O2. INDICATION IS THAT SIEVE BED FILTERS (BOTTOM) ARE NOT SEATED PROPERLY ALLOWING SIEVE MATERIAL TO BECOME PULVERIZED AND BYPASS FILTER. SIEVE DUST IS MIGRATING INTO THE MANIFOLD VALVE AND MUFFLER. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
650577 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other