FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4171238 · Received October 14, 2014

Report

Report Number
3004753838-2014-29420
Event Type
Malfunction
Date Received
October 14, 2014
Date of Event
August 3, 2014
Report Date
August 5, 2014
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RETURNED COMPLAINT DEVICE WAS VISUALLY INSPECTED AND FOUND NO OBSERVATIONS RELATED TO THE COMPLAINT. A REVIEW OF THE RECEIVER DATA LOG CONFIRMED A HARDWARE ERROR CODE. THIS COMPLAINT WAS DEEMED REPORTABLE UPON COMPLETION OF DEVICE EVALUATION ON 09/18/2014. THE ROOT CAUSE WAS DETERMINED TO BE A HARDWARE COMPONENT FAILURE.

Description of Event or Problem · 1

PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO REPORT THAT ON (B)(6) 2014 PATIENT'S RECEIVER WOULD NOT TURN ON. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
651123 G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT22430-BLU

Patients

Seq Age Sex Outcome Treatment
1 7 YR