SMALL BATTERY DRIVE
Report
- Report Number
- 8030965-2014-10543
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Date of Event
- September 29, 2014
- Report Date
- September 29, 2014
- Manufacturer
- SYNTHES OBERDORF
- Product Code
- HWE
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
THIS DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITIONS WERE CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS MOST LIKELY DUE TO WORN COMPONENTS AND DETERIORATION OF BALL BEARINGS DUE TO NORMAL USE AND WEAR OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS DISCOVERED THAT THE SMALL BATTERY DRIVE DEVICE HAD LOUD NOISES AND THE TRIGGERS WERE STICKY ON THE DEVICE. THE EVENTS WERE NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 651121 | SMALL BATTERY DRIVE | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT | HWE | SYNTHES OBERDORF | 003661 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |