FDA Adverse Event Malfunction Summary report: N

ECHELON LINEAR CUTTER

MDR report key: 4171161 · Received October 14, 2014

Report

Report Number
3005075853-2014-07069
Event Type
Malfunction
Date Received
October 14, 2014
Date of Event
September 24, 2014
Report Date
September 24, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K110385
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS FOUND THAT ONE PLE60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH AN ECR60T CARTRIDGE LOADED IN THE DEVICE. ADDITIONALLY, THE RELOAD WAS RECEIVED WITH THE RIGHT SIDE FULLY FIRED, LEFT SIDE OUTER ROW FULLY FIRED AND THE LEFT TWO INNER ROWS PARTIALLY FIRED 1/3. UPON EVALUATION OF THE RELOAD, THE CARTRIDGE BODY, ONE PIECE SLED AND SOME DRIVERS WERE NOTED TO BE DAMAGED. NO OBVIOUS DAMAGE TO THE CARTRIDGE DECK WAS NOTED, WHICH SUGGESTS THE CARTRIDGE, MAY NOT HAVE BEEN FIRED OVER A HARD OBJECT. AS ADDITIONAL TESTING, THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GASTRIC SLEEVE PROCEDURE, ON THE INITIAL FIRING WITH A BLACK LOAD, THE STAPLES DID NOT FORM ALL OF THE WAY. THERE WERE NO PATIENT CONSEQUENCES. CASE COMPLETED BY OVER SEWING THE STAPLE LINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
649516 ECHELON LINEAR CUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA L4F16K

Patients

Seq Age Sex Outcome Treatment
1 RELOAD-ECR60T, LOT # L4F23Z