ECHELON LINEAR CUTTER
Report
- Report Number
- 3005075853-2014-07069
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Date of Event
- September 24, 2014
- Report Date
- September 24, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS FOUND THAT ONE PLE60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH AN ECR60T CARTRIDGE LOADED IN THE DEVICE. ADDITIONALLY, THE RELOAD WAS RECEIVED WITH THE RIGHT SIDE FULLY FIRED, LEFT SIDE OUTER ROW FULLY FIRED AND THE LEFT TWO INNER ROWS PARTIALLY FIRED 1/3. UPON EVALUATION OF THE RELOAD, THE CARTRIDGE BODY, ONE PIECE SLED AND SOME DRIVERS WERE NOTED TO BE DAMAGED. NO OBVIOUS DAMAGE TO THE CARTRIDGE DECK WAS NOTED, WHICH SUGGESTS THE CARTRIDGE, MAY NOT HAVE BEEN FIRED OVER A HARD OBJECT. AS ADDITIONAL TESTING, THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4).
IT WAS REPORTED THAT DURING A GASTRIC SLEEVE PROCEDURE, ON THE INITIAL FIRING WITH A BLACK LOAD, THE STAPLES DID NOT FORM ALL OF THE WAY. THERE WERE NO PATIENT CONSEQUENCES. CASE COMPLETED BY OVER SEWING THE STAPLE LINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 649516 | ECHELON LINEAR CUTTER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | L4F16K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RELOAD-ECR60T, LOT # L4F23Z |