PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-23739
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Date of Event
- August 3, 2014
- Report Date
- August 17, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP PASSED DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33 AND EXCESSIVE NO DELIVERY TESTS. NO UNEXPECTED EXCESSIVE NO DELIVERY ALARM NOTED. THE INSULIN PUMP WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHED LCD WINDOW AND SCRATCHED RESERVOIR TUBE WINDOW.
IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP WAS ALARMING NO DELIVERY. CUSTOMER STATES THAT THEY HAVE CHANGED THE RESERVOIR AND INFUSION SET, YET ISSUE HAS CONTINUED. CUSTOMER VERIFIED CONNECTION BETWEEN THE RESERVOIR AND INFUSION SET WAS SECURE. DURING TROUBLESHOOTING, CUSTOMER WAS ASKED TO PERFORM A MANUAL PRIME OF INSULIN. CUSTOMER STATED THE INSULIN DID EXIT. NEXT, CUSTOMER WAS ASKED TO PERFORM A FIXED PRIME. CUSTOMER STATED DEVICE ALARMED NO DELIVERY. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 138 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 651188 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |