INRATIO PT/INR TEST STRIPS
Report
- Report Number
- 2027969-2014-00658
- Event Type
- Malfunction
- Date Received
- July 14, 2014
- Date of Event
- June 24, 2014
- Report Date
- June 24, 2014
- Manufacturer
- ALERE SAN DIEGO, INC
- Product Code
- GJS
- PMA / PMN Number
- K072987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
INVESTIGATION/CONCLUSION: IT IS INDICATED THAT PRODUCT IS NOT RETURNING FOR EVAL. THEREFORE, INVESTIGATION OF THE COMPLAINT TO DETERMINE ROOT CAUSE CANNOT BE COMPLETED. SINCE THE PRODUCT ASSOCIATED WITH THE COMPLAINT WAS NOT RETURNED, A REVIEW OF IN-HOUSE TESTING DATA WAS PERFORMED. RETAIN STRIP TESTING RESULTS MET BOTH ACCURACY AND REPEATABILITY CRITERIA. THE PRODUCT PERFORMED AS EXPECTED AND NO PRODUCT DEFICIENCIES WERE OBSERVED. ALTHOUGH THE ROOT CAUSE ANALYSIS DID NOT INCLUDE RETURN TESTING, IMPROPER TECHNIQUE WAS IDENTIFIED IN THE COMPLAINT. THIS COULD NOT BE RULED OUT AS A CAUSE OF THE UNEXPECTED RESULTS. THE MFG RECORDS FOR THE LOT WERE REVIEWED AND THERE WERE NO ISSUES RELATED TO THIS COMPLAINT. NO FURTHER INVESTIGATION WILL BE PURSUED AT THIS TIME. BASED ON THE INFO AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT INRATIO INR RESULTS. RESULTS ARE AS FOLLOWS: DATE: (B)(6) 2014; INRATIO INR: 1.5, 3.9, 4.4, AND 3.6. THERAPEUTIC RANGE: 2.0-2.9. THE TESTING WAS PERFORMED WITHIN FIFTEEN (15) - TWENTY (20) MINUTES. REPORTEDLY THE FINGER WAS TOUCHED TO THE SAMPLE WELL. THERE WAS NO REPORTED ADVERSE PT SEQUELA. THERE WAS NO ADD'L INFO PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 410581 | INRATIO PT/INR TEST STRIPS | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC | 100071 | 033862 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | COUMADIN 5 MG 3X WEEKLY AND 2.5 MG 4X WEEKLY,| LEVOQUIN (RECENTLY COMPLETED)| INRATIO MONITOR SN UNK, |