FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4170988
·
Received October 14, 2014
Report
- Report Number
- 1031452-2014-13935
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Report Date
- September 24, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER INDEPENDENT REPAIR CENTER STATEMENT THE UNIT HAD LOW O2 OR YELLOW LIGHT. THE KEY FAILURE WAS THE NYLON TIES FOR THE PE VALVE WAS LEAKING. ADDITIONAL MALFUNCTIONS INCLUDE THE SIEVE BEDS WERE SATURATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 650832 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |