FDA Adverse Event Malfunction Summary report: N

11CM ANGLE ATTACHMENT

MDR report key: 4170946 · Received October 14, 2014

Report

Report Number
1045834-2014-13974
Event Type
Malfunction
Date Received
October 14, 2014
Report Date
May 15, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES (B)(4) AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE OR UNCHANGED. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR.

Additional Manufacturer Narrative · 1

DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE DEVICE HAD MULTIPLE PARTS REQUIRING REPLACEMENT. ABOVE-SPECIFICATION TEMPERATURE WAS ALSO FOUND. THIS WAS ATTRIBUTED TO IMPROPER MAINTENANCE. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPLACED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FOR AN UNSPECIFIED REASON. THIS IS REPORT 1 OF 1 FOR COM-(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
649902 11CM ANGLE ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1