PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4170851
·
Received October 14, 2014
Report
- Report Number
- 2032227-2014-37871
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Date of Event
- September 15, 2014
- Report Date
- September 15, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
Additional Manufacturer Narrative · 1
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
Description of Event or Problem · 1
IT WAS REPORTED THE CUSTOMER HAD RECURRING MOTOR ERROR ALARMS. THE CUSTOMER STATED THEY DID NOT DROP OR BUMP THEIR INSULIN PUMP. THE INSULIN PUMP WAS ALSO NOT EXPOSED TO A HIGH MAGNETIC FIELD. CUSTOMER'S BLOOD GLUCOSE WAS 154 MG/DL. NO ADDITIONAL INFORMATION PROVIDED.