FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4170833 · Received October 14, 2014

Report

Report Number
1031452-2014-13774
Event Type
Malfunction
Date Received
October 14, 2014
Report Date
September 24, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

PER INDEPENDENT REPAIR CENTER STATEMENT, THE UNIT HAD LOW O2. THE KEY FAILURE WAS THE SIEVE BEDS HAD LOW O2. ADDITIONAL MALFUNCTIONS INCLUDE THE PILOT VALVE WAS ALARMING, AND THE 4-WAY VALVE WAS LEAKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
649452 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other