FDA Adverse Event Injury Summary report: N

SHARPS 3 GALLON

MDR report key: 4170550 · Received October 2, 2014

Report

Report Number
3007508552-2014-00002
Event Type
Injury
Date Received
October 2, 2014
Date of Event
September 8, 2014
Report Date
December 12, 2018
Manufacturer
SHARPS COMPLIANCE, INC.
Product Code
MMK
PMA / PMN Number
K083129
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION OF ACTUAL CONTAINER NOT POSSIBLE DUE TO CONTAINER NOT RETURNED TO MANUFACTURER. AT SUCH A TIME AS SHARPS RECEIVES THE CONTAINER SHARPS WILL PERFORM AN INVESTIGATION, INCLUDING BUT NOT LIMITED TO THE WALL THICKNESS OF THE SECTION OF THE CONTAINER WHERE THE NEEDLE IS PROTRUDING. ALL SHARPS CONTAINERS ARE PUNCTURE RESISTANT, NOT PUNCTURE PROOF. PRODUCT LABELING ON THE CONTAINER STATES "CONTAINERS ARE PUNCTURE RESISTANT, NOT NECESSARILY PUNCTURE PROOF", AND INSTRUCTIONS FOR USE STATE "DO NOT OVERFILL (FILL LINE IS NOTED ON CONTAINER LABEL). LID MUST FIT DOWN TIGHTLY. BASED ON THE REVIEW OF PREVIOUS NEEDLE PROTRUDING FROM CONTAINER REPORTS IT IS USUALLY THE CONTAINER BEING OVERFILLED THAT LEADS TO NEEDLE PROTRUSIONS.

Description of Event or Problem · 1

ON (B)(6) 2014 AT 9 AM, (B)(6), (WITH (B)(6)), CALLED IN TO REPORT AN INCIDENT. CUSTOMER WAS CARRYING SHARPS CONTAINER BACK TO PHARMACY AND ACCIDENTLY BUMPED LEG INTO BOTTOM OF CONTAINER AND CUSTOMER WAS POKED BY NEEDLE THAT WAS PROTRUDING OUT OF THE BOTTOM OF THE CONTAINER. CONTAINER HAS BEEN PLACED IN A SECURE AREA SO NO ONE CAN ACCESS IT. CUSTOMER DID SEE A DOCTOR AND SHE HAS TAKEN A BLOOD TEST. SHE IS CURRENTLY AWAITING RESULTS OF THIS TEST TO MAKE SURE NO INFECTION OR DISEASE WAS INJECTED INTO HER SYSTEM. PER A1 WE WILL SEND OUT CUSTOMER ITEM 12801 - 28 GALLON MEDICAL PROFESSIONAL SHARPS RECOVERY SYSTEM SO THAT THE CONTAMINATED CONTAINER CAN BE SENT BACK FOR QUALITY INSPECTION. AS OF (B)(6) 2014, THE CONTAINER HAS NOT BEEN RETURNED TO SHARPS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
616761 SHARPS 3 GALLON SHARPS CONTAINER MMK SHARPS COMPLIANCE, INC. 13000 UNK

Patients

Seq Age Sex Outcome Treatment
1 Other