FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 4170545 · Received October 14, 2014

Report

Report Number
2245578-2014-00075
Event Type
Malfunction
Date Received
October 14, 2014
Date of Event
September 24, 2014
Report Date
September 24, 2014
Manufacturer
ABBOTT POINT OF CARE
Product Code
MMI
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION WAS COMPLETED ON 11/10/2014. RETAIN AND RETURNED CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(6) 2014, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 0.17 (FALSE POSITIVE) ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD:I-STAT, TEST DATE: (B)(6) 2014, DRAW TIME: 12:35, RESULT: 0.00, SAMPLE: A; I-STAT, (B)(6) 2014, 14:51, 0.17, B; I-STAT, (B)(6) 2014, 15:03, 0.00, C; ARCHITECT, (B)(6) 2014, 15:03, 0.02, C; THERE WERE NO INJURIES REPORTED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
651666 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE NA U14176

Patients

Seq Age Sex Outcome Treatment
1