I-STAT TROPONIN (CTNI) CARTRIDGE
MDR report key: 4170545
·
Received October 14, 2014
Report
- Report Number
- 2245578-2014-00075
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Date of Event
- September 24, 2014
- Report Date
- September 24, 2014
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- MMI
- PMA / PMN Number
- K031739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE INVESTIGATION WAS COMPLETED ON 11/10/2014. RETAIN AND RETURNED CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
ON (B)(6) 2014, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT RESULT OF 0.17 (FALSE POSITIVE) ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD:I-STAT, TEST DATE: (B)(6) 2014, DRAW TIME: 12:35, RESULT: 0.00, SAMPLE: A; I-STAT, (B)(6) 2014, 14:51, 0.17, B; I-STAT, (B)(6) 2014, 15:03, 0.00, C; ARCHITECT, (B)(6) 2014, 15:03, 0.02, C; THERE WERE NO INJURIES REPORTED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.