ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-29138
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Report Date
- October 1, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 12/22/2014 - DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/03/2014 WITH THE FOLLOWING FINDINGS: THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THE BATTERY COMPARTMENT WAS CRACKED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(4) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THE REPORTER ALLEGED THAT THE DISPLAY SCREEN COULD NOT BE READ SAFELY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 651108 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR |