FDA Adverse Event
Malfunction
Summary report: N
8CM ANGLE ATTACHMENT, BLACK MAX
MDR report key: 4170502
·
Received October 14, 2014
Report
- Report Number
- 1045834-2014-14056
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Report Date
- May 15, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBE
- PMA / PMN Number
- PK831756
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE CONTROL UNIT DID NOT OPERATE. THIS WAS ATTRIBUTED TO NORMAL WEAR. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER. (B)(6).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE BEARING AND GEAR WERE DAMAGED. PER SERVICE REPORT, HEAT, VIBRATION, BONDING SURFACE LOOSE. NOSE TUBE AND SCREW LOOSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 651058 | 8CM ANGLE ATTACHMENT, BLACK MAX | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |