FDA Adverse Event Malfunction Summary report: N

8CM ANGLE ATTACHMENT, BLACK MAX

MDR report key: 4170502 · Received October 14, 2014

Report

Report Number
1045834-2014-14056
Event Type
Malfunction
Date Received
October 14, 2014
Report Date
May 15, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBE
PMA / PMN Number
PK831756
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE CONTROL UNIT DID NOT OPERATE. THIS WAS ATTRIBUTED TO NORMAL WEAR. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER. (B)(6).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE BEARING AND GEAR WERE DAMAGED. PER SERVICE REPORT, HEAT, VIBRATION, BONDING SURFACE LOOSE. NOSE TUBE AND SCREW LOOSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
651058 8CM ANGLE ATTACHMENT, BLACK MAX DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1