SEDLINE MODULE PATIENT CABLE
Report
- Report Number
- 2031172-2014-00207
- Event Type
- Death
- Date Received
- September 24, 2014
- Date of Event
- August 25, 2014
- Report Date
- August 25, 2014
- Manufacturer
- MASIMO CORPORATION
- Product Code
- GWQ
- PMA / PMN Number
- K140188
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NURSE
- Health Professional
- N
Narratives
ATTEMPTS FOR ADDITIONAL INFO HAVE BEEN MADE. IT HAS BEEN INDICATED THAT THE PT CABLE COULD BE RETURNED. AS OF THIS REPORT, THE DEVICE HAS NOT BEEN RECEIVED BY MASIMO TO ALLOW FOR AN ANALYSIS TO BE PERFORMED. IF NEW INFO IS OBTAINED, A FOLLOW-UP REPORT WILL BE SUBMITTED.
CUSTOMER REPORTED THAT "JUST WANTED YOU TO KNOW THAT I HAD A PT WITH THIS READ OUT ON HER SEDLINE. ISOELECTRIC EEG (SEE PIC BELOW), PRESSED ON THE DOTS AND GOT A REACTION THROUGH THE CABLE TO THE DEVICE, ALL THE DOTS WERE GREEN ON THE MONITOR DEMONSTRATING GOOD CONTACT, SO WE WERE THINKING SHE HAD A BAD PROBLEM. I CHANGED THE SWEEP SPEED AND VOLTAGE JUST TO BE CERTAIN THAT WASN'T AN ISSUE, BUT THE READOUT REALLY DIDN'T CHANGE. AFTER DISCUSSION WITH DR. RAMSAY AND OUR PHARMACIST WE STARTED TITRATING DOWN HER BENZOS AND FENTANYL. RETURNED TO SEE THE PT SEVERAL HOURS LATER AND EEG READOUT WAS UNCHANGED. I DISCUSSED THE CASE WITH OUR INTENSIVIST AND HE PUT IN AN ORDER FOR A FULL EEG. WHILE WAITING, OUT OF CURIOSITY, I CHANGED THE BOX AND THE CABLE AND STARTED GETTING THIS READOUT. SWITCHED THE NEW CABLE TO THE OLD BOX, AND THE EEG CONTINUED TO READ CORRECTLY, NO LONGER ISOELECTRIC. THIS IS NOT THE FIRST TIME I HAVE HAD THIS ISSUE WITH A CABLE THAT IS NOT PICKING UP THE EEG CORRECTLY, YET WE ARE STILL GETTING PSI AND DSA. THERE DOESN'T APPEAR TO BE ANY WARNINGS THAT THERE IS A CABLE ISSUE. I HAVE THE CABLE ON MY DESK. PLEASE ADVISE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 595987 | SEDLINE MODULE PATIENT CABLE | OXIMETER | GWQ | MASIMO CORPORATION | 24295 | ||
| 596213 | SEDLINE MODULE PATIENT CABLE | OXIMETER | GWQ | MASIMO CORPORATION | 24295 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male | Death |