FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4170460 · Received October 14, 2014

Report

Report Number
2032227-2014-37973
Event Type
Malfunction
Date Received
October 14, 2014
Date of Event
September 12, 2014
Report Date
September 15, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED DISPLACEMENT TEST, A21 ERROR TEST, PRIME/A33 TEST AND EXCESSIVE NO DELIVERY TEST. NO EXCESSIVE NO DELIVERY ALARM AND NO A21 ALARM NOTED. THE INSULIN PUMP WAS RECEIVED CRACKED CASE AT DISPLAY WINDOW CORNER, WITH CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE CUSTOMER RECEIVED AN EXCESSIVE NO DELIVERY ALARM. THERE WAS ALSO SEVERAL EXCESSIVE NO DELIVERY ALARMS IN THE ALARM HISTORY. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
651022 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWP

Patients

Seq Age Sex Outcome Treatment
1