FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4170444 · Received October 14, 2014

Report

Report Number
2032227-2014-37793
Event Type
Malfunction
Date Received
October 14, 2014
Date of Event
September 15, 2014
Report Date
September 15, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. THE INSULIN PUMP HAS CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP WAS ALARMING BUTTON ERROR. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LED TO THE ALARM OR IF THE DEVICE WAS DROPPED, BUT MAY HAVE BEEN EXPOSED TO PERSPIRATION. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 166 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.