FDA Adverse Event Malfunction Summary report: N

11 CM ANGLE ATTACHMENT

MDR report key: 4170410 · Received October 14, 2014

Report

Report Number
1045834-2014-13894
Event Type
Malfunction
Date Received
October 14, 2014
Report Date
February 4, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE DEVICE WAS DAMAGED. THIS WAS ATTRIBUTED TO NORMAL WEAR. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, TECHNICIAN FOUND THAT THE DEVICE HAD VIBRATIONS AND THE BALL BEARING IN THE EXTENSION TUBE WAS DEFECTIVE. THIS IS REPORT 1 OF 1 FOR (B)(4).