FDA Adverse Event Injury Summary report: N

ESPOCAN COMBINED SPINAL/EPIDURAL TRAYS

MDR report key: 4170271 · Received September 26, 2014

Report

Report Number
2523676-2014-00346
Event Type
Injury
Date Received
September 26, 2014
Date of Event
August 25, 2014
Report Date
August 28, 2014
Manufacturer
B. BRAUN MEDICAL, INC.
Product Code
CAZ
PMA / PMN Number
K813186
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT HAS BEEN IDENTIFIED AS B BRAUN MEDICAL INC. (B)(4). ONE USED CATHETER AND ONE UNUSED CATHETER WERE RECEIVED FOR EVAL. THE UNUSED CATHETER APPEARED INTACT AND NO ANOMALIES OR FRACTURED ENDS WERE VISUALLY OBSERVED. THE USED CATHETER WAS MISSING THE BLUE TIP AND APPEARED FRACTURED AT THE TIP END. UNDER MICROSCOPIC OBSERVATION, THE FRACTURED END APPEARED STRETCHED AND FRAYED. THIS TYPE OF DAMAGE IS CONSISTENT WITH EPIDURAL CATHETERS IN WHICH A SECTION IS PULLED BEYOND ITS DESIGN CAPABILITIES. THIS CAN OCCUR WHEN A CATHETER BECOMES LODGED BETWEEN RIGID BODY STRUCTURES. REVIEW OF THE DISCREPANCY MANAGEMENT SYSTEM DATABASE PERFORMED FOR THE REPORTED LOT NUMBER DID NOT REVEAL ANY ABNORMALITIES OR NONCONFORMANCES OF THIS NATURE. NO ADVERSE QUALITY TRENDS OF THIS NATURE WERE IDENTIFIED DURING THE COMPLAINT REVIEW PROCESS FOR THE REPORTED CATALOG NUMBER OR CATHETER MATERIAL NUMBER. THERE WERE NO OTHER REPORTS OF THIS NATURE AGAINST THE REPORTED LOT NUMBER. ALL AVAILABLE INFO HAS BEEN FORWARDED TO THE DEVICE MANUFACTURER OF THE CATHETER. IF ADD'L PERTINENT INFO BECOMES AVAILABLE, A F/U REPORT WILL BE FILED.

Description of Event or Problem · 1

AS REPORTED BY THE USER FACILITY: REPORTS WHEN THE CATHETER WAS REMOVED THE TIP BROKE OFF IN THE PT. REPORTER IS UNAWARE IF THERE WERE ANY PROBLEMS WITH THE INSERTION OR REMOVAL. THE CATHETER DID NOT LOOK STRETCHED. THE TIP WAS NOT SURGICALLY REMOVED AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
599767 ESPOCAN COMBINED SPINAL/EPIDURAL TRAYS PERIFIX 20 GA. EPIDURAL CATHETER CAZ B. BRAUN MEDICAL, INC. ES1827KDS 0061354255

Patients

Seq Age Sex Outcome Treatment
1 UNK Other