FDA Adverse Event
Malfunction
Summary report: N
COULTER LH 750 HEMATOLOGY ANALYZER
MDR report key: 4170238
·
Received October 14, 2014
Report
- Report Number
- 1061932-2014-02592
- Event Type
- Malfunction
- Date Received
- October 14, 2014
- Date of Event
- September 22, 2014
- Report Date
- September 22, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ON (B)(4) 2014, A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE ANALYZER AND REPLACED LV14B; IT WAS BROKEN AT THE BASE. THE FSE ALSO REPLACED VL 14A, IT WAS BROKEN ALSO. (B)(4).
Description of Event or Problem · 1
A BECKMAN COULTER FIELD SERVICE ENGINEER IDENTIFIED A BROKEN VALVE WHILE SERVICING A COULTER LH 750 HEMATOLOGY ANALYZER IN A CUSTOMER'S FACILITY ON (B)(4) 2014. THERE WAS NO REPORT OF ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.