FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4170121 · Received October 14, 2014

Report

Report Number
2032227-2014-37738
Event Type
Malfunction
Date Received
October 14, 2014
Date of Event
September 15, 2014
Report Date
September 15, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP ALARMED DURING BASIC OCCLUSION TEST DUE TO PROTRUDED AND LOOSE DRIVE SUPPORT DISK. INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND CRACKED END CAP.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT INSULIN PUMP IS STUCK IN THE PRIMING LOOP PHASE. CUSTOMER STATED THAT INSULIN PUMP DID NOT RECEIVE THE 2ND SERIES OF BEEPS NOR DID NUMBERS APPEAR ON THE SCREEN. CUSTOMER'S BLOOD GLUCOSE READING WAS 169 MG/DL. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.