FDA Adverse Event Malfunction Summary report: N

MINILINK TRANSMITTER

MDR report key: 4170112 · Received October 14, 2014

Report

Report Number
2032227-2014-37773
Event Type
Malfunction
Date Received
October 14, 2014
Date of Event
August 18, 2014
Report Date
September 18, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE TRANSMITTER WAS UNABLE TO CHARGE DUE TO A BROKEN COW CATCHER AND DAMAGE TO ALL CONTACT PINS. THE FUNCTIONAL TESTS COULD NOT BE PERFORMED DUE TO THE CHARGE ANOMALY.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE TRANSMITTER DID NOT BLINK WHEN CONNECTED TO ANY CHARGER OR DURING THE TEST PLUG PROCEDURE. THE BLOOD GLUCOSE READING WAS 206 MG/DL. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
649289 MINILINK TRANSMITTER CGM MDS MEDTRONIC MINIMED MMT-7703NA

Patients

Seq Age Sex Outcome Treatment
1 65 YR