FDA Adverse Event Injury Summary report: N

PROMOTE RF CRT-D

MDR report key: 4170054 · Received January 13, 2014

Report

Report Number
2938836-2014-02916
Event Type
Injury
Date Received
January 13, 2014
Date of Event
November 17, 2011
Manufacturer
ST. JUDE MEDICAL INC., CRMD
Product Code
NIK
PMA / PMN Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.

Description of Event or Problem · 1

IT WAS REPORTED THAT T-WAVE OVERSENSING RESULTED IN 7 INAPPROPRIATE SHOCKS. THIS ISSUE WAS CLINICALLY RESOLVED BY REPROGRAMMING THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
34014 PROMOTE RF CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR, NIK NIK ST. JUDE MEDICAL INC., CRMD 3213-36

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention