FDA Adverse Event Other Summary report: N

STRATAFIX KNOTLESS TISSUE CONTROL DEVICE

MDR report key: 4168976 · Received October 7, 2014

Report

Report Number
3008845715-2014-00058
Event Type
Other
Date Received
October 7, 2014
Date of Event
August 20, 2014
Report Date
October 6, 2014
Manufacturer
SURGICAL SPECIALTIES PUERTO RICO INC.
Product Code
GAM
PMA / PMN Number
K072028
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

SPITTING/INFLAMMATION/REACTION. METHOD: THE ACTUAL PRODUCT INVOLVED WILL NOT BE RETURNED. PER ELECTRONIC COMMUNICATION FROM THE CUSTOMER IT WAS STATED THAT THERE ARE THREE POSSIBLE LOTS INVOLVED, THE THREE SUSPECT LOTS PROVIDED ARE; MBNQ080, MBZK810, MBKB350. RELEVANT PORTIONS OF THE DEVICE HISTORY RECORDS WERE REVIEWED. FINISHED GOOD PRODUCTS WERE RECEIVED INTO INVENTORY WITHOUT QUALITY ISSUES. THE PRODUCT FROM THESE FINISHED GOOD LOTS AND ALL OF THE COMPONENTS MET SURGICAL SPECIALITIES PUERTO RICO INC. REQUIREMENTS THROUGHOUT THE INCOMING, MANUFACTURING AND THE FINAL INSPECTION PROCESSES. NO INVENTORY AVAILABLE FOR THE SUSPECT FINISHED GOOD LOTS. INFLAMMATORY TISSUE REACTION, IS A KNOWN RISK WITH ANY SUTURE MATERIAL. THE MOST PROBABLE ROOT CAUSE FOR POST OPERATIVE INFLAMMATION/REACTION CAN NOT BE DETERMINED WITH CERTAINTY BASED ON THE INFORMATION AND WITHOUT REVIEWING THE ACTUAL DEVICE INVOLVED OR TESTING STERILE SAMPLES FROM THE SAME FINISHED GOODS LOT. (B)(4). ITEM NUMBER AND FINISHED GOOD LOT NUMBER ARE UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED BY THE FACILITY CONTACT, "WE'VE HAD 5 PATIENT'S COMPLAIN OF SUTURES SPITTING OUT AND HAVING IRRITATION AROUND THE INCISION. ALSO NOTICING THE WOUND OPENING UP. "NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. REF. PR COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
626229 STRATAFIX KNOTLESS TISSUE CONTROL DEVICE BARBED MATERIAL/NEEDLES GAM SURGICAL SPECIALTIES PUERTO RICO INC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 63 YR NONE MADE AVAILABLE