FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4163661 · Received October 10, 2014

Report

Report Number
2032227-2014-19273
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
September 11, 2014
Report Date
September 11, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO UNEXPECTED NUMBER RAMPING OR SCROLLING DURING TESTING. THE INSULIN PUMP HAS CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW, STAINED END CAP STICKER AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP'S KEYPAD WAS NOT RESPONDING DURING BOLUS. CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THIS ERROR. BLOOD GLUCOSE LEVEL WAS 54 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
639286 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAB

Patients

Seq Age Sex Outcome Treatment
1 39 YR