FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR

MDR report key: 4163651 · Received October 10, 2014

Report

Report Number
3015876-2014-01196
Event Type
Malfunction
Date Received
October 10, 2014
Date of Event
September 15, 2014
Report Date
September 15, 2014
Manufacturer
PHYSIO-CONTROL, INC
Product Code
LDD
PMA / PMN Number
K063119
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EXAMINED THE CUSTOMER'S DEVICE AND VERIFIED THE REPORTED ISSUE. PHYSIO THEN REPLACED BOTH THE SYSTEM CONTROLLER (SC) AND USER INTERFACE (UI) PCB ASSEMBLIES AND AFTER OBSERVING PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING THE UNIT WAS RETURNED TO THE CUSTOMER FOR USE.

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EXAMINED THE REMOVED SYSTEM CONTROLLER PCB ASSEMBLY AND DETERMINED THAT THE CAUSE OF THE REPORTED ISSUE WAS AN INTEGRATED CIRCUIT (IC) CHIP, DESIGNATOR U61. THE REMOVED USER INTERFACE PCB ASSEMBLY WAS AN ANCILLARY PART AND THERE WAS NO FAILURE OF THE ASSEMBLY.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT WHEN ATTEMPTING TO POWER ON THEIR DEVICE IT LOCKED UP WITH A WHITE DISPLAY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
640378 LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) LDD PHYSIO-CONTROL, INC 20

Patients

Seq Age Sex Outcome Treatment
1